Service personnel must wear calibrated radiation monitoring badges to track cumulative exposure.
Attach a fluorescent alignment tool or a specialized alignment phantom to the sensor holder. Take a low-dose test exposure.
The best source for the most up-to-date documentation. kodak 8000 service manual
The unit uses three Class 2 laser beams for patient positioning. Service checks should ensure these beams are properly calibrated and not "stared into". 3. Common Troubleshooting Steps
The is a variant of the panoramic system with cephalometric capabilities. Documentation for this model is more readily available: The best source for the most up-to-date documentation
Vertical movement failure due to a jammed actuator or a faulty lift motor.
This series includes several models. Knowing your exact model is essential: kodak 8000 service manual
Analyze the resulting image. If the beam center is offset from the sensor centerline, manually adjust the collimator blades using the micro-metric adjustment screws located on the X-ray tube head. Sensor Gain and Dark Current Calibration
Understanding the physical and electronic layout of the Kodak 8000 is critical before performing any service tasks. The manual divides the unit into several interconnected subsystems: